| Model | G4-1040/80 |
|---|---|
| Ultrasonic frequencies | 40 & 80 kHz |
| Rated power | 1,000 W |
| Power adjustment | 10–100 %Linear control across the range |
| Power supply | 220 VAC ±10 %, 50/60 Hz |
| Control topology | Full-bridge phase shift |
| Power regulation | Constant power control |
| Frequency control | Automatic trackingLoad self-adaptation |
| Frequency switching | 24 VDC ±10 %, 20 mA |
| Communications | RS485 (Modbus RTU) |
| Protection class | IP21 |
| Ambient conditions | 0–40 °C, 0–90 % RH |
| Enclosure size | 412 × 160 × 238 mmL × W × H — 19-inch cabinet integration supported |
| Conversion efficiency | Up to 95 % |
| Full-bridge phase shift | Stable, efficient power conversion for industrial ultrasonic loads. |
|---|---|
| Constant power output | Holds controlled output as the load and the bath conditions change. |
| Automatic tracking | Follows the resonant frequency and adapts to impedance changes — no manual retuning visit. |
| Peak power | Available on demand for stubborn contamination — the figure is confirmed against the model. |
| Sweep | Breaks up standing waves and evens out the energy distribution. |
| Monitoring | Voltage, current, phase, status, temperature, run time and both frequencies on the front LCD. |
| Protection | Over-current, over-voltage and short circuit, internally. |
| Integration | Remote control to PLC and HMI, RS485 Modbus RTU, 19-inch cabinet mounting. |
One generator, two selected operating frequencies — aggressive soil removal and fine cleaning on the same platform. Pick the frequency pair your parts need, then the power your transducer set draws.
Which frequency should I choose? → How loud is each band? →
We do not sell a generator on its own to drive transducers we have not specified, and we do not sell transducers to run on a generator we did not supply. It is not a policy about paperwork — the two are tuned to each other, and an unmatched pair usually means no cavitation, a drive running hot, and both halves replaced inside a year.
There is no warranty on an unmatched pair. If you already own one half, tell us what it is and roughly when it went in and we will tell you where you stand before you spend anything.
Model codes read power then frequencies: G4‑10 40/80 is 1,000 W at 40 and 80 kHz. Tap a row to load that model into the configurator above.
| Model | Frequency | Rated power |
|---|
These nine are the dual-frequency range, and 1,500 W is the ceiling for running two bands from one generator. Built as a single-frequency unit the same platform drives higher power, because all of it goes into one band instead of being shared across two. Tell us the transducer set and we will confirm the generator, or the number of them, that drives it.
A generator is tuned to the transducers it drives. Four things have to line up: the frequency the generator tracks, the total capacitance of the array wired to it, the impedance and phase that array presents once it is in the liquid, and the voltage across the group under load. We specify all four against each other, so the pair is quoted, built and commissioned as one item.
A new generator will not rescue an old set. Transducers age — output falls away, the resonant point drifts and the impedance climbs — so an older array is no longer the load a new drive was tuned for. A set of mixed ages is worse again: the weakest element pulls the rest off tune and the drive spends its power hunting for a resonance that has moved. What you get is a generator running hot or dropping out, a bath cleaning unevenly, and transducers failing earlier than they should — and from the outside there is no way to tell which end is at fault, so you end up replacing both anyway.
There is no warranty on an unmatched pair. Equipment warranty covers a generator and transducers supplied and matched by Misonics. A Misonics transducer driven by a generator we did not supply, or a Misonics generator fitted to transducers we have not specified, is not covered — we have no way to characterise the combination, so we will not stand behind it. If you already own one half, tell us what it is and roughly when it went in, and we will tell you where you stand before you spend anything.
Tell us what it is driving — tank size, transducer count and the frequency your parts need — and we will confirm the model and quote it. If you already know the model you want, say so and we will just quote it.